WIKA TR12-M Benutzerhandbuch

Seite 24

Advertising
background image

GB

8. Safety-related instructions ... / 9. Calculation examples ...

11419904.02 09/2013 GB/D

WIKA operating instructions models TR12, TC12 (Ex d)

24

8.3 Variant 3:

ATEX/IECEx Ex d certified temperature transmitters.

The evaluation is carried out over a current (4 … 20 mA), voltage (0 … 10 V) or field bus

signal, which is generated by an ATEX/IECEx Ex d certified temperature transmitter.

Use in zone 1, marking II 2G Ex d IIC Gb

The flameproof housing or the connection head is in zone 1 (or zone 2). The sensor is in

zone 1. In the case of a separation of Ex-zones, a thermowell (from corrosion-resistant

steel, min. wall thickness 1 mm) must be used.

The main markings for models TR12-B and TC12-B are found on the certified connection

housing or Ex d field transmitter.

The TR12-M and TC12-M modules are marked through a foil plate on the neck tube.

For a possible use at the partition to zone 0 with a thermowell, the approvals and conditions

of the relevant Ex d field transmitters must be followed.

Temperature class classification, ambient temperatures

For the measuring insert with neck tube, the following ambient temperature range applies:

(-50)

1)

-40 … +85 °C

1) The value in brackets applies only for specific low-temperature designs

The permissible ambient temperatures of the certified Ex d field transmitters must be taken

from the respective operating instructions or data sheets. Through deviations in temperature

ranges, limitations can occur.

8.4 Use in methane atmospheres

As a result of the higher Minimum Experimental Safe Gap (MESG) and Minimum Ignition

Current (MIC) of methane, the instrument can also be used in hazardous gas atmospheres

caused by this.

9. Calculation examples for self-heating at the thermowell tip

The self-heating at the thermowell tip depends upon the sensor type (TC/RTD), the

measuring insert diameter and the thermowell design. The table below shows the possible

combinations. The heating at the sensor tip of the bare measuring insert is clearly higher;

the representation of these values was omitted on the grounds of the required assembly

with a thermowell. The table shows that when a failure occurs, thermocouples produce

much less self-heating than resistance thermometers.

Advertising
Dieses Handbuch ist für die folgenden Produkte bezogen werden: