WIKA operating instructions, model CPH65I0

14061681.02 03/2013 GB/D

GB

2. Safety

WARNING!

Do not use the calibrator if it is damaged. Before using the

instrument, check that there are no cracks or missing plastic parts on

the case. Pay particular attention to the insulation of the connectors.

The battery cover must be closed and locked in place before the

instrument is operated.

Remove the test cables from the instrument before opening the

battery compartment.

Inspect the test cables for damaged insulation or exposed metal.

Check the continuity of the leads. Damaged test leads should be

replaced before using the instrument.

When using test probes, keep fingers away from the test probe

contacts. Place fingers behind the finger guards on the test probes.

Do not use the instrument if it is not working properly. The instrument

protection might be compromised. If in doubt, have the instrument

checked.

Only use the instrument in classified areas for which the calibrator is

approved.

Disconnect test leads before changing to another measurement or

generation function.

To avoid false readings, replace the battery as soon as the battery

indicator appears.

Use the correct connection, the proper function and the correct

measuring range for measurements.

When screwing an adaptor onto the NPT thread of the calibrator, a

sealing material must also be used (e.g. PTFE tape).

When screwing on, use a spanner on the pressure connection for

fixing to the calibrator, in order to avoid unnecessary stress on the

case.