Gb 6. commissioning, operation – WIKA CPH65I0 Benutzerhandbuch

Seite 28

Advertising
background image

28

WIKA operating instructions, model CPH65I0

14061681.02 03/2013 GB/D

GB

6. Commissioning, operation

6.4.6 Reduced screen resolution

Due to the high accuracy of the CPH65I0, the measured value is displayed with a higher

resolution. In some cases, this is a disadvantage. For this, the CPH65I0 has a function

for lowering the display resolution. This function removes the last digit.

In order to switch the function on or off, perform the following:

With the calibrator switched on, press the F3 key in order to activate the "MORE"

option. Press the "NEXT" key until "RESOLUTION" appears in the left-hand text field.

Then press the F1 key to activate the function.

Select "ON" or "OFF" in order to switch the lower display resolution on or off.

Use "DONE" in order to return to the home menu.

6.4.7 Setting the resistance thermometer probe

Use the "PROBE TYPE" option of the 'RTD Probe Type' main menu to access the

menu for resistance thermometer probe selection.

There are four types of resistance thermometers: "P100-385", "P100-392",

"P100-JIS" and "CUSTOM". With "CUSTOM", custom resistance thermometers

can be selected. Use the "SELECT" option to select the desired probe type and the

"DONE" option to store the change and return home.

Advertising