3. Specifications

3. Specifications

Sensor technology

Measuring range

mbar

1)

-25 ... +25 -70 … +70 -350 ... +350 -500 … +500

Overpressure limit

mbar

70

200

700

1,000

Accuracy

% FS

0.1

0.05

0.035

Measuring range

bar

-1 ... +1

1)

-1 ... +2

1)

0 … 1

1)

0 ... 2

1)

Overpressure limit

bar

2

4

2

4

Accuracy

% FS

0.025

Measuring range

bar

0 ... 3.5

2)

0 ... 7

2)

0 ... 10

2)

0 ... 20

2)

Overpressure limit

bar

13

13

13

40

Accuracy

% FS

0.025

Measuring range

bar

0 ... 35

2)

0 ... 70

0 ... 100

0 ... 200

Overpressure limit

bar

70

200

200

400

Accuracy

% FS

0.025

Measuring range

bar

0 ... 350

0 ... 700

Overpressure limit

bar

700

1,000

Accuracy

% FS

0.035

Measuring range

bar abs.

0 ... 1

0 ... 2

0 ... 7

0 ... 10

0 ... 20

Overpressure limit

bar abs.

2

4

13

13

40

Accuracy

% FS

0.025

Measuring range

mbar diff.

1) 3)

0 ... 25

0 ... 70

0 ... 350

Overpressure limit

mbar diff.

70

200

700

Accuracy

% FS

0.1

0.05

0.035

Measuring range

mbar diff.

1) 3)

0 ... 2,000 0 ... 3,500 0 ... 7,000

Overpressure limit

mbar diff.

4,000

7,000

10,000

Accuracy

% FS

0.025

Type of pressure

Relative pressure, absolute pressure, vacuum and differential pressure

4)

Pressure connection

⅛ NPT female (incl. adapter ⅛ NPT male to G ½ B male)

5)

Pressure medium

all liquids and gases which are compatible with 316 SS stainless steel

1)

Resolution

5-digit

Current

Measuring range

0 … 24 mA

Resolution

1 µA

Accuracy

0.015 % of measured value ±2 µA

Temperature

Measuring range

-40 … +150 °C

Resolution

0.01 °C

Accuracy

0.015 % of measured value ±20 mΩ, or 0.2 °C for complete measuring chain

(Pt100 resistance thermometer and CPH65I0)

1) Non isolated: Only use the pressure measuring ranges marked in this way with clean and non-corrosive gases.

2) Pressure measurement possible in vacuum range to -1 bar.

3) For differential pressure sensors with a measuring range of 25 mbar, the maximum static pressure is limited to 70 mbar. For the

measuring ranges 70, 350, 2,000, 3,500 and 7,000 mbar the static pressure is limited to a maximum of 10 bar.

4) The differential pressure sensor is only possible with the CPH65I0-S1 (1-channel version). Both pressure connections for the

differential pressure measurement are located at the bottom of the calibrator.

5) Adapter not included in delivery for North America.