2. Safety

Connection values

Max. voltage

Uo = DC 7.14 V

Max. current

Io = 1.12 mA

Max. power

Po = 2 mW

Max. effective internal capacitance Co = 240 µF
Max. effective internal inductance

Lo = 1 H

Power supply circuit

Max. voltage

Ui = DC 30 V

Max. current

Ii = 80 mA

Max. power

Pi = 750 mW

Max. effective internal capacitance Ci = 0 nF
Max. effective internal inductance

Li = 0 mH

WARNING!

Further hazardous area safety instructions:

Observe the operating information and the relevant state regulations

concerning use in hazardous areas (e.g. EN IEC 60079-14).

The model CPH65I0 intrinsically-safe hand-held pressure calibrator has

been designed for use in hazardous areas. In these areas there is the

possibility of inflammable or explosive gases being present.

The model CPH65I0 intrinsically-safe hand-held pressure calibrator has

been designed to be intrinsically safe. This means that connection with

other equipment that is in an intrinsically-safe circuit will not lead to arcs

capable of causing ignition, so long as the parameters are complied with.

Information/Approval for hazardous locations

WARNING!

Hazardous Areas

A hazardous area, as used in these operating instructions, refers to

an area made hazardous by the potential presence of flammable or

explosive vapours. These areas are also referred to as hazardous

locations.