Gb 6. commissioning, operation – WIKA CPH65I0 Benutzerhandbuch

Seite 24

Advertising
background image

24

WIKA operating instructions, model CPH65I0

14061681.02 03/2013 GB/D

GB

6. Commissioning, operation

6.3.1 Internal relative pressure sensor

When a relative pressure sensor is selected in the {active display} and the ZERO key

is pressed, the calibrator subtracts the current displayed reading from the output value.

The zero limits are within 5 % of the full scale range of the selected sensor. If the display

indicates "OL" the zero function will not operate.

6.3.2 Internal absolute pressure sensor

If an absolute pressure sensor is selected in the {active display} and the ZERO button

is pressed, the calibrator instructs the user to enter the reference pressure value with

"SET" or to reset with "DEFAULT". The reference instrument should be at least four

times as accurate as the absolute pressure sensor of the CPH65I0.

If, alternatively, "DEFAULT" is selected, the reversion to the factory setting must be

accepted with "CONFIRM" or the action aborted with "CANCEL".

If "SET" is selected, the calibrator prompts the user to enter the barometric reference

pressure. The arrow keys, F2 and F3, can be used for this. The sensor connection

should be open (vented) to atmosphere while performing this procedure.