6. Commissioning, operation

To perform a switch test, the following steps must be taken:

1. Switch to SETUP 4 (standard switch test).

SETUP 4: The upper display is set to [P1] ST, all other displays are switched off.

The pressure switch test can be performed with the following functions

[P1] ST or [P2] ST.

2. Connect the calibrator to the pressure switch using the switch terminals. The polarity

of the terminals is not important. Then connect the calibrator and the pump to the

pressure switch.

6.10 Performing a pressure switch test