Faults, Faults causes measures – WIKA CPP1200-X Benutzerhandbuch

Seite 33

Advertising
background image

33

WIKA operating instructions model CPP1200-X

GB

14076968.01 12/2013 GB/D

8. Faults

Faults

Causes

Measures

Equipment provides

pressure but pressure

decays when valves A

and B are operated.

Incorrect operating procedure

being used.

Ensure that correct operating

procedure is being followed (see

chapter 6.6).

If unable to locate a cause.

Return test pump to

DH-Budenberg/WIKA for

investigation.

Equipment provides

pressure but pressure

decays to lower value

then remains steady.

Insufficient liquid in pressure

balance.

Check liquid level in reservoir. Fill

reservoir with correct liquid as

necessary (see chapter 6.5.1).

Air in the system

Pre-fill component under test with

appropriate liquid. If necessary

re-fill pressure balance with

appropriate liquid.

If unable to locate a cause.

Return test pump to

DH-Budenberg/WIKA for

investigation.

Internal damage

Return test pump to

DH-Budenberg/WIKA for

investigation.

Incorrect operating procedure

being used.

Ensure that correct operating

procedure is being followed (see

chapter 6.6).

If unable to locate a cause.

Return test pump to

DH-Budenberg/WIKA for

investigation.

The spindle pump of

the test pump becomes

very stiff to operate

when the test pump is

being used in the range

below 140 bar (2,000 lb/

in²).

Internal damage

Return test pump to

DH-Budenberg/WIKA for

investigation.

The spindle pump of

the test pump becomes

very stiff to operate

when the test pump is

being used in the range

over 140 bar (2,000 lb/

in²).

Incorrect operating procedure

being used.

Ensure that correct operating

procedure is being followed (see

chapter 6.6).

If unable to locate a cause.

Return test pump to

DH-Budenberg/WIKA for

investigation.