WIKA operating instructions model CPP1200-X

GB

14076968.01 12/2013 GB/D

2. Safety

2.1 Intended use

Test pumps serve as pressure generators for the testing, adjustment and calibration

of mechanical and electronic pressure measuring instruments through comparative

measurements. These pressure tests can take place in the laboratory or workshop, or

on site at the measuring point.

The comparison test pump features two connections, for test item and reference

measuring instrument, which can be used in any order. If one connects the test item

and a sufficiently accurate reference measuring instrument to the test pump, on

actuating the pump, the same pressure will act on both instruments. By comparison of

the two measured values at any given pressure value, a check of the accuracy and/or

adjustment of the pressure measuring instrument under test can be carried out.

The integrated dual-area spindle pump enables rapid filling of the test system and

smooth pressure generation up to 1,200 bar. At the same time, the precise, adjustable

spindle pump also enables fine pressure adjustment.

A control schematic for pressure generation on the instrument base facilitates quick

and easy operation. The pump is further characterised by the spindle which runs

solely within the pump body. This eliminates any harmful bending moment from an

externally running spindle and, especially for field operation, there is the advantage that

dimensions of this pump do not alter during operation through the rotation of the spindle.

The term “reference measuring instrument” in these operating

instructions refers to any pressure measuring instrument, such as:

pressure gauge, electrical pressure measuring instrument and pressure

transmitter with electrical output. The comparison test pump system

is only as accurate as the reference measuring instrument used. The

reference measuring instrument should be regularly calibrated in order to

ensure that its accuracy is maintained.

The instrument has been designed and built solely for the intended use described here,

and may only be used accordingly.

The technical specifications contained in these operating instructions must be

observed. Improper handling or operation of the instrument outside of its technical

specifications requires the instrument to be taken out of service immediately and

inspected by an authorised DH-Budenberg/WIKA service engineer.

Handle mechanical precision measuring instruments with the required care (protect

from humidity, impacts, strong magnetic fields, static electricity and extreme

temperatures, do not insert any objects into the instrument or its openings).