INSTALLMION AND SAFETY INSTRUCTIONS PM 5414 V

1.1.3

Line Voltage Setting and Fuses

Before plugging in the power cable, make sure that the instrument is set to the correct line voltage.

WARNING

Changing fuses and modifying power cables to local power must be done by quali­

fied service personnel who are aware of the hazards involved.

On delivery from the factory the instrument is set to one of the following line voltages.

Instrument

version

Instrument

code no.

Line voltage

setting

Deiivered

power cabie

PM 5414 V

9452 054 140x1

220 V

Universal Europe

PM 5414 V

9452 054 140x3

120 V

North America

PM 5414 V

9452 054 140x4

240 V

United Kingdom

PM 5414 V

9452 054 140x5

220 V

Switzerland

PM 5414 V

9452 054 140x8

240 V

Australia

The line voltage setting and the corresponding fuse specification are indicated on the rear panel.

Make sure that replacement fuses are of the type and current rating specified type. The use of repaired

fuses, and/or the short circuiting of the fuse holders are prohibited. Do not defeat this important safety

feature.

The instrument can be set to the following line voltages: 100 V, 120 V, 220 V and 240 V a.c. These

nominal voltages can be selected by means of the voltage selector, located on the rear panel next

to the line voltage connector. The fuse is located In a holder at the same place. For line voltage selec­

tion or replacement of the fuse, remove the power cable and pry open the compartment with a small

screwdriver (see illustration).