Line Voltage Setting and Fuses
Before plugging in the power cable, make sure that the instrument is set to the correct line voltage.
WARNING
Changing fuses and modifying power cables to local power must be done by quali
fied service personnel who are aware of the hazards involved.
On delivery from the factory the instrument is set to one of the following line voltages.
Instrument
version
Instrument
code no.
Line voltage
setting
Deiivered
power cabie
PM 5414 V
9452 054 140x1
220 V
Universal Europe
PM 5414 V
9452 054 140x3
120 V
North America
PM 5414 V
9452 054 140x4
240 V
United Kingdom
PM 5414 V
9452 054 140x5
220 V
Switzerland
PM 5414 V
9452 054 140x8
240 V
Australia
The line voltage setting and the corresponding fuse specification are indicated on the rear panel.
Make sure that replacement fuses are of the type and current rating specified type. The use of repaired
fuses, and/or the short circuiting of the fuse holders are prohibited. Do not defeat this important safety
feature.
The instrument can be set to the following line voltages: 100 V, 120 V, 220 V and 240 V a.c. These
nominal voltages can be selected by means of the voltage selector, located on the rear panel next
to the line voltage connector. The fuse is located In a holder at the same place. For line voltage selec
tion or replacement of the fuse, remove the power cable and pry open the compartment with a small
screwdriver (see illustration).