OPERATING INSTRUCTIONS PM 5414 V

3.4.2

General Functional Test

After POWER ON, the instrument is automatically set to the operating mode to which it was set before

power off (test patterns).

-

Under the PATTERN area on the front panel, select patterns GREYSCALE/ COLOR BAR/MULTI-

BURST.

-

Connect the VIDEO OUTPUT at the front panel or the scart-connector atthe rear to a suitable CTV

or monitor (for example correct TV system, video input voltage; if necessary, adjust amplitude by

turning VIDEO AMPL).

- Check the correct video reproduction on the screen.

- Select and check additional test patterns.

- Connect PM 5414 V VIDEO OUTPUT to an oscilloscope (75 Q termination).

- Select patterns GREYSCALE/WHITE.

- Set the VIDEO AMPLITUDE to stop position 1 V.

- Check that the video amplitude is 1 V (peak-peak), accuracy <5 %.