II Technical Information

II - 8 Er

ro

r Messag

es

DRO Error Message

Explanation

Error:

The ARCSIN function can only

operate on values from - 1 to 1.

Range error on the value to fetch the arcsin of.

Error:

The TAN function is undefined at

90, and -90 degrees.

Range error on the value to fetch the tangent of.

Error:

The square root of a negative

number is undefined.

Can not take the square root of a negative number.

Error:

Divide by zero error.

Can not divide by zero.

Unable to transfer configuration

parameters. The external device is not

ready.

The external device hooked to the serial port is not working. Check the physical

connection, and verify that the serial communications parameters are correct.

Error:

One or more of the configuration

parameters was invalid, and could not be

loaded.

The configuration file being imported is corrupt. Import file again. If error persists,

import a different configuration file.

Error:

Parameter was invalid. It has been

set to it’s previous value.

An invalid configuration parameter was discovered on import, and was reset to the

previous value. Check the configuration file that failed, and import again.

Error:

The valid range for scale factor is

-10 to -0.1, and 0.1 to 10.

The current scale factor value is out of range. Adjust the scale factor value to

reflect a valid number.

Error:

The valid range for the error

correction factor in parts per million is

-99999 to +99999.

Out of range for the error correction factor defined within error compensation.

When defining the error correction factor, the user must use the correct range.

Error:

Interpolation Error.

The G38 chip corresponding to the encoder input failed verification.