Menu guidance of the user interface – WIKA TIF52 Benutzerhandbuch
Seite 26

Function
Factory setting
on/off
Activating or deactivating the alarm function;
if a value exceeds or falls below a set alarm limit, the display shows the warning
symbol and the measured value starts flashing.
Setting range:
- OFF
- ON
OFF
Low alarm
Setting the value which releases the alarm function when this value is not reached.
Setting range:
Start value of the indication range up to the set value of the high alarm
4,000
High alarm
Setting the value upon exceeding of which the alarm function is released.
Setting range:
Set value of the low alarm up to the end value of the indication range
20,000
8.2.2 Submenu "Min/Max memory"
Function
Factory setting
Delete min/max
Function for deleting the maximal value memory
Pressing the SELECT key twice deletes the maximum value memory.
dEL
min/max on/off
Activation of the min./max. display;
Iif the min./max. display is switched on, the display switches cyclically between the
current measured value (display time 5 s), the minimum value and the maximum
value (display time 2 s).
For the display of the maximum values, the unit on the display is replaced by min.
or max.
Setting range:
- OFF
- ON
OFF
26
WIKA operating instructions models TIF50, TIF52
GB
14009686.02 03/2012 GB/D
8.2.1 Submenu "Alarm"
8. Menu guidance of the user interface