Power supply

DC 10 ... 35 V

Max. permissible residual ripple

10 % at 24 V / maximum 300 Ω Load

Power supply input

protected against reverse polarity

Power supply effect

± 0.025 % / V

Electromagnetic compatibility

(EMC)

2004/108/EC, DIN EN 61326 emission (Group 1, Class B) and immunity

(industrial application)

5)

, and also per NAMUR NE21

Temperature units

configurable °C, °F, K

Info data

TAG No., descriptor and message can be stored in transmitter

Configuration and calibration data permanently stored in EEPROM
Electrical connection

Angular DIN connector Form A, M12 x 1 circular connector, 4-pin

Explosion protection (option)

Intrinsically safe to Ex-i (ATEX) gas/dust, per Directive 94/9/EC

Marking: II 1G Ex ia IIC T* or II 2D Ex iaD 21 T*

Safety-related maximum values for the current loop (+ and - connections):
Ui = DC 30 V

C

i

= 6.2 nF I

i

= 120 mA L

i

= 110 µH Pi = 800 mW

Readings in % refer to the measuring span

For a correct determination of the overall measuring error, both sensor and transmitter measuring deviations have to

be considered.

1) For detailed specifications for Pt100 sensors, see Technical Information IN 00.17 at www.wika.com.

2) The temperature transmitter should therefore be protected from temperatures over 85 °C

3) For measuring spans smaller than 50 K additional 0.1 K

4) ± 0.2 % for measuring ranges with a lower limit less than 0 °C

5) Use RTDs with shielded cable, and ground the shield on at least one end of the lead, if the cables are longer than 30 m or leave the

building.

Load diagram

3. Specifications

The permissible load depends on

the loop supply voltage.

Voltage U

B

in V

Load R

A

in

Ω

0

10

24

30

36

1182

909
636

For setting the measuring range see chapter "7. Configuration of model TR30-W".