Safety-related instructions for ex operation – WIKA TR30 Benutzerhandbuch

Seite 23

Advertising
background image

14016442.01 06/2011 GB/D

WIKA operating instructions model TR30 per ATEX

23

GB

Example for model TR30-P-A

Resistance thermometer, diameter: 6 mm

T

amb

= 40 °C

Supplied power = 50 mW

[Alternative example: 500 mW]

Temperature class for T3, Instrument Category 1

Additional information

Temperature class for T3 = 200 °C

Safety factor for type-tested instruments (for T6 to T3)

1)

= 5 K

Safety factor for applications for Instrument Category 1 (Zone 0)

2)

= 80 %

1) EN 50014: 1997 Section 23.4.6.1

2) EN 1127-1: 1997 Section 6.4.2

To determine: max. permissible temperature in °C at the sensor tip T

M

Thermal resistance [R

th

in K/W] from Table 4 = 75 K/W

Supplied power [P

o

in W] = 50 mW = 50 *10

-3

W = 0.05 W

[0.5 W]

Self-heating [in K] = 75 K/W * 0.05 W = 3.75 K

[37.5 K]

T

M

= (temperature class for T3 - safety factor for type-examined instruments) x safety factor

for applications for instrument category 1 - self-heating [in K]

T

M

= (200 °C – 5 K) x 0.8 – 3.75 K

[(200 °C – 5 K) x 0.8 – 37.5 K]

T

M

= 152.25 °C

[118.5 °C]

The permissible temperature in °C at the sensor tip (TM) = 152.25 °C [118.5 °C]

9. Safety-related instructions for Ex operation