6 Commissioning, Operation
14 WIKA Operating Instruction, Model CPH6300
14043054
.01
07
/20
12
G
B/
D
Menu
Parameter Value
Description
or
Set Configuration: General configurations
Unit: Select measuring unit
*,**
Arrow to bar,
mbar …
Measuring value is displayed in corresponding unit, the
selectable units depend on connected sensor
Arrow to
“user”
Measuring value is displayed in freely adjustable user-unit
bar, mbar
Unit “user”: base unit
0000, 000.0
… .0000
Unit “user”: decimal point setting (only if user-unit selected)
-19999 ...
19999
Unit “user”: multiplication factor
Sea Level: Sea-level correction
*,**
oFF
Sea-level correction off
on
Sea-level correction on (see6.4.3)
-2000 ... 9999 Sea-level in [m]
*,**
Rate: Measuring rate
*
Slo
Slow measuring rate (4 Hz filtered, low power consumption)
FASt
Fast measuring rate, filtered (1000 Hz)
P.dEt
Peak detection: fast measuring rate, unfiltered (1000 Hz)
Averaging Filter
*
1 … 120
Averaging period in seconds
oFF
Averaging function is deactivated
Auto Power-Off: Select power-off delay
1 ... 120
Power-off delay in minutes.
Device will be automatically switched off as soon as this time has
elapsed if no key is pressed/no interface communication takes
place.
oFF
Automatic power-off function deactivated (continuous operation)
Background illumination
(factory settings: 5 s)
oFF
Illumination deactivated
5 … 120
Turn off illumination after 5 … 120s
on
Illumination always on