6 Commissioning, Operation
16 WIKA Operating Instruction, Model CPH6300
14043054
.01
07
/20
12
G
B/
D
Menu
Parameter
Value
Description
or
Set Output: Universal output adjustments
oFF
Interface and analog output off
SEr:
Serial interface activated
dAC:
Analog output activated
01,11..91
Base address for serial interface communication
(only at Out = SEr)
CH 1, CH 2 or
DIF
CH
Channel that should be used for analog output
(only at Out = dAC)
***
sensor dep.
e.g.-5.00..5.00 mbar
Zero-point setting for Out = dAC: Measuring value which
should correspond to output 0 V
sensor dep.
e.g.-5.00..5.00 mbar
Slope setting for Out = dAC: Measuring value which should
correspond to output 1 V
Set Corr: Input adjustment (1-channel version)
*
Zero adjustment/offset of sensors
oFF
No zero adjustment of sensors
sensor dep.
e.g.-5.00..5.00 mbar
The offset of sensor will be displaced by this value to
compensate for deviations of the probe or of the measuring
device.
Slope adjustment of sensors
oFF
No slope adjustment of sensor
-2.000 ... 2.000
The slope of sensor will be changed by this factor [%] to
compensate deviations of probe or measuring device
Set Corr: Input adjustment (2-channel version)
*
***
Zero adjustment/offset of sensors 1
oFF
No zero adjustment of sensors 1
sensor dep.
e.g.-5.00..5.00 mbar
The offset of sensor 1 will be displaced by this value to
compensate for deviations of the probe or of the measuring
device.
Slope adjustment of sensors 1
oFF
No slope adjustment of sensor 1
-2.000 ... 2.000
The slope of sensor 1 will be changed by this factor [%] to
compensate deviations of probe or measuring device
Zero adjustment/offset of sensors 2
oFF
No zero adjustment of sensors 2
sensor dep.
e.g.-5.00..5.00 mbar
The offset of sensor 2 will be displaced by this value to
compensate for deviations of the probe or of the measuring
device.
Slope adjustment of sensors 2
oFF
No slope adjustment of sensor 2
-2.000 … 2.000
The slope of sensor 2 will be changed by this factor [%] to
compensate deviations of probe or measuring device
GB