Hand-held thermometer cth 6200, Scal, Conf unit – WIKA CTH6200 Benutzerhandbuch

Seite 11: Offs, P.off, Al.lo, Al.hi, Logg func, Cycl, Cloc

Advertising
background image

WIKA Betriebsanleitung Hand-Held Thermometer

2146872 09/2004 GB/D

11

GB

WIKA Operating Instructions Hand-Held Thermometer

Hand-Held Thermometer
CTH 6200

Menu

Param.

Values

Meaning

„Menu“

or

SEt

Set Configuration: Generic Settings

ConF

Unit

°C:

All temperatures values are in degrees Celsius

*

°F:

All temperatures values are in degrees Fahrenheit

rES

0.1 °:

Resolution 0.1 °C

*

0.01°

Resolution 0.01 °C

*

Auto:

Resolution is selected automatically

Lin

E.751

Measuring / sensor curve according to EN 60751

*

USEr

In preparation

OFFS

-2.50 °C ... 2.50 °C

The zero point of the measurement will be displaced by this value to

*

resp. -4.50 °F ... 4.50 °F

compensate for deviations in the probe or in the measuring device.

oFF:

Zero displacement inactive (=0.0 °C)

SCAL

-2.000 ... 2.000

The scale of the measuring will be changed by this factor [in %] to

*

compensate deviations of temperature probe or measuring device.

oFF:

Scale correction factor inactive ( = 0.000)

P.oFF

1 ... 120

Auto Power-Off delay in minutes.

Device will be automatically switched off as soon as this time has

elapsed if no key is pressed / no interface communication takes place.

oFF

Auto Power-off function inactive (continuous operation)

Out

oFF:

Function of the output: No output function, lowest power consumption

SEr:

Output is serial interface

dAC:

In preparation

Adr.

01, 11 .. 91

In preparation

SEt

Set Alarm: Settings of alarm function

AL.

AL.

on

Alarm Sensor 1 on, with horn-sound

no. So

Alarm Sensor 1 on, without horn-sound

oFF

No alarm function

AL.Lo

min range ... AL.Hi

Min alarm rail (not when AL.oFF)

AL.Hi

AL.Lo ... max range

Max alarm rail (not when AL.oFF)

SEt

Set Logger: Configuration of logger function

*

LoGG

Func

CYCL

Cyclic: logger function ,cyclic logger`

*

Stor

Store: logger function ,individual value logger`

*

oFF

No logger function

*

CYCL

0:01 ... 60:00

Cycle time of cyclic logger [minutes:seconds]

*

SEt

Set Clock: Setting of real time clock

CLOC

CLOC

HH:MM

Clock: setting ot time hours:minutes

dAtE

TT.MM

Date: setting of time day.month

YEAr

YYYY

Year: setting of time year

Advertising