WIKA Betriebsanleitung Hand-Held Thermometer

2146872 09/2004 GB/D

13

GB

WIKA Operating Instructions Hand-Held Thermometer

Hand-Held Thermometer
CTH 6200

The sensor curve following EN60751 uses the international
temperature scale ITS90 and following formulas:

Temperatures <0 °C:

R

neg

(T): = 100 [1 + 3,908310

-3

* T - 5,77510

-7

* T

2

- 4,18310

-12

* (T - 100) * T

3

]

Temperatures >=0 °C:
R

pos

(T): = 100 (1 + 3,908310

-3

* T - 5,77510

-7

* T

2

)

2.4 (OFFs) Zero point correction / Offset

A zero point correction can be carried out for the measured
temperature:

temperature displayed = temperature measured - offset

Standard setting: ´off´ = 0.0 °, i.e. no zero point correction will be

carried out. The zero point correction is mainly used to compen-
sate for sensor deviations. Unless ´off´ is set, this value will be

displayed shortly after the device is switched on;
during operation it will be identified by means of the offset arrow in

the display.

2.5 (Scal) Scale correction

The scale of the measuring can be influenced by this setting
(factor is in %).

displayed temperature [°C] = measured temperature [°C] * (1 + Scal/100)

or respectively

displayed temperature [°F] = (measured temperature [°C] - 32

°F) * (1 + Scal/100) + 32 °F

Standard setting: ´off´ =0.000, i.e. temperature is nor corrected.
Together with the zero displacement (see above) this factor is

mainly used to compensate for sensor deviations.
Unless the factor is set to ´off´, the Corr arrow in the display shows

an active scale correction.