Hand-held thermometer cth 6200, Fault and system messages – WIKA CTH6200 Benutzerhandbuch

Seite 19

Advertising
background image

WIKA Betriebsanleitung Hand-Held Thermometer

2146872 09/2004 GB/D

19

GB

WIKA Operating Instructions Hand-Held Thermometer

Hand-Held Thermometer
CTH 6200

6. Fault and system messages

Display

Meaning

Remedy

Replace battery

Check power supply, replace it when
necessary

Replace battery

Check power supply, replace it when
necessary

Replace battery

Netzgerät überprüfen / austauschen

Disconnect battery or power supply,
wait some time, re-connect

Return to manufacturer for repair

Connect sensor to socket

Return to manufacturer for repair

Is the value exceeding the measuring
range specified?
-> measuring value to high!

Check probe

Return to manufacturer for repair

Check: is the value below the
measuring range specified?
-> measuring value to low!

Check probe

Return to manufacturer for repair

-> Set resolution to 0.1 ° or Auto

-> Set resolution to 0.1 ° or Auto

Return to manufacturer for repair

Do not leave the specified range

Low battery voltage, device will coninue
to work for a short time

If mains operation:
wrong voltage

Battery empty

Mains operation without battery:
wrong voltage

Battery empty

Mains operation without battery:
wrong voltage

System error

Device defective

Sensorerror: son sensor connected

Sesnor / cable or device defective

Value exceeding measuring range

Wrong probe connected

Sensor / cable or device defective

Value below display range

Wrong probe connected

Sensor / cable or device defective

Value exceeding display range

Value below display range

System error

Out of range

No display
or strange
symbols,
device is not
responding
to input

----

Err.1

Err.2

Err.3

Err.4

Err.7