7. Calibration services

DKD-certificates - other certificates:

If device should be certificated for its accuracy, it is the best
solution to return it with the referring sensors to the manufacturer.

Only the manufacturer is capable to do efficient recalibration if
necessary to get results of highest accuracy!

8. Specifications

Characteristics of the complete measuring chain (digital instrument and probe):
Input

1

Sensor type

Pt 100 (4-wire)

Measuring range ex stock

-50 ... +250 °C

Total measurement

1)

0.2 K

uncertainty

1)

(On request a higher accuracy can be reached by an adjustment at a
certain temperature range.)

Calibration *

Calibration report 3.1.B according to DIN EN 10204

Display

2 x 4 1/2-digit LCD display for actual value and additional information

Resolution

0.01 K up to 200 °C; than 0.1 K

Units

°C, °F

CTH 6200 Digital Instrument specific details:
Measuring range

-99.99 ... +99.99 °C or -199.9 ... +650.0 °C respectively
(dependent on probe)

Functions via key press

Min-, Max-memory, Hold, Logger (Start/Stopp)

Functions via menu

Min-, Max-alarm (acoustic/visual), autom. Power-Off-Function
[1 ... 120 min], Zero point / span correction, Logger settings

Data logger

- individual value logger: up to 99 recordings incl. time via keypress
- cyclic logger: automatic recording up to 16,200 values incl. time,
with cyclic time selectable from 1 ... 3600 seconds

Interface (serial)

RS 232 interface via interface converter (USB adapter on request)

Power supply

9 V-zinc-carbon battery (incl. in scope of delivery),
{alternative: 9 V rechargeable battery or mains supply}

Power consumption

approx. 1.25 mA

Nominal temperature

25 °C

Temperature drift

0.002 °C/K

Permissible ambient

-10 ... +50 °C

temperature
Air humidity

0 ... 95 % relative humidity without moisture condensation

Storage temperature

-25 ... +70 °C

Housing

impact-resistant ABS, membrane keyboard, transparent panel

Weight

approx. 160 g

CTP 62x0 Temperature sensor specific details (Pt 100):
Probes available

Immersion and penetration probe

Electrical connection

4-pole Mini-DIN plug with PVC cable (approx. 1.0 m)

Temperature range
- Immersion probe

CTP 6210

-50 ... +250 °C

- Penetration probe

CTP 6290

-50 ... +250 °C

Material wetted parts

Immersion and penetration probe: stainless steel

Dimensions

Sensor length: 300 mm; hand grip: 135 mm
Sensor diameter: D = 3 mm

1) The total measurement uncertainty is stated for the whole measuring chain after adjustment and

calibration.

*)

Calibration at 23 °C ± 2 °C ambient temperature.

{}

Items in curved brackets are optional extras for additional price.