WIKA Operating Instructions Hand-Held Thermometer

Hand-Held Thermometer
CTH 6200

4. General information about temperature sensors

Heat loss caused by probe construction:

Especially when measuring temperatures which deviate very
much from the ambient temperature, measuring errors often

occur if the heat loss caused by the probe is not considered.
When measuring fluids therefore the probe should be emerged

sufficiently deep and be stirred continuously. When measuring
gases the probe should also emerge as deep as possible in the

gas to be measured (e.g. when measuring in channel/pipes) and
the gas should flow around the probe at sufficient flow.

The tube length should be selected long enough, that temperature

keeps low at the handle.

Self Heating
The measuring current of the instrument is just 0.3 mA. Be-
cause of this comparably low current practically no self heating
effect has to be considered, even at air with low movement

the

self heating is <= 0.01 °C.

Cooling by Evaporation
When measuring air temperature the probe has to be dry.

Otherwise the cold due to the evaporation causes too low
measurings.

4.1 Available sensors

The device is designed to be connected to all sensors of the

CTH6200 series (see valid price list CTH6200).

5. The serial interface

By means of the serial interface and a suitable electrically isolated
interface adapter the device can be connected to a computer for

data transfer (USB-adapter on request), to visualise the data with
the WIKA Datalogger evaluation software GSOFT (see valid price

list CTH6200). To avoid transmission errors, there are several

security checks implemented e.g. CRC.