Konfiguration (lds2010-kompatibel) – INFICON LDS3000 Modular Leak Detector Benutzerhandbuch

Seite 24

Advertising
background image

24

Betrieb

6.2.4

Konfiguration (LDS2010-kompatibel)

LDS2010
setting
menu item
22

Analog
output
channel

Function LDS2010

Function LDS3000

Scale for
leakrate

upper limit
(10 V = ...)

1

1

Leak rate Mantissa in used unit
1 ... 10 V

Leak rate mantissa

don't care don't care

1

2

Leak rate Exponent (step function)
in used unit
1 ... 10 V, 0,5 V/decade,
1 V = 1E-12

Leak rate exponent

don't care don't care

2

1

Leak rate log. in used unit
1 ... 10 V, 0,5 V/decade,
1 V = 1E-12

Leak rate log.

0,5 V/dec. 1E6 [used unit]

2

2

Pressure p1 log. in used unit
1 ... 10 V, 0,5 V/decade,
1 V = 1E-3 mbar

Pressure p1

don't care don't care

3

1

Leak rate matissa in mbar·l/s
1 ... 10 V

Leak rate mantissa

don't care don't care

3

2

Leak rate exponent (step function)
in mbar·l/s
1 ... 10 V, -1 V/decade,
0 V = 1E0 mbar·l/s

LR exponent invers

don't care don't care

4

1

Leak rate log.
0 ... 10 V, 1 V/decade,
0 V = 1E-10 mbar·l/s

Leak rate log.

1 V/dec.

1,00E+00

4

2

Pressure p1 log. in mbar
1 V/decade, 2,5 ... 8,5 V,
2,5 V = 1E-3 mbar,
5,5 V = 1E0 mbar

p1 1 V/dec.

don't care don't care

5

1

Leak rate mantissa in used unit
1 ... 10 V rise, 0,7 ... 10 V fall

LR mantissa hyst.

don't care don't care

5

2

Leak rate exponent in used unit
1 ... 10 V, 0,5 V/decade,
0 V = 1E-14

Leak rate exponent

don't care don't care

6

1

Leak rate log. in Pa·m³/s
0 10 V, 1 V/decade,
0 V = 1E-12 Pa·m³/s =
1E-12 mbar·l/s

Leak rate log.

1 V/dec.

1E-2 mbar·l/s

6

2

Pressure p1 log. in Pa
1 V/decade, 2,5 ... 8,5 V,
2,5 V = 1E-3 mbar

p1 1 V/dec.

don't care don't care

8

1

Leak rate log. in Pa·m³/s
0 ... 10 V, 1 V/decade,
0 V = 1E-12 Pa·m

3

/s =

1E-12 mbar·l/s

Leak rate log.

1 V/dec.

1E-2 mbar·l/s

8

2

Pressure p2 log. in Pa
1 V/decade, 2,5 ... 8,5 V,
2,5 V = 1E-3 mbar

p2 1 V/dec.

don't care don't care