Configure[1..n]:acquisition:mode – EXFO FTB-7000 OTDR Series for FTB-500 Benutzerhandbuch
Seite 396

SCPI-Befehlsreferenz
382
Serie FTB-7000
Product-Specific Commands—Description
:CONFigure[1..n]:ACQuisition:MODE
Description
This command specifies the mode that will be
used for the next acquisition.
Acquisition: Allows the OTDR to perform a
standard acquisition.
Auto Setting: Lets the OTDR evaluates the length
of the fiber and finds the appropriate range and
pulse width.
Check First Connector: Used to detect a low
injection level.
Real Time: Used to view sudden changes in the
fiber under test. In this mode, measurements are
not allowed.
*RST sets the current acquisition mode to
ACQUISITION.
Syntax
:CONFigure[1..n]:ACQuisition:MODE<wsp>ACQ
uisition|ASETting|CFConnector|REAltime
Parameter(s)
Mode:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
ACQuisition|ASETting|CFConnector|REAltime.