SCPI-Befehlsreferenz

396

Serie FTB-7000

Product-Specific Commands—Description

:CONFigure[1..n]:ANAlysis:

IORefraction

Description

This command sets the index of refraction that

will be used for the next acquisition.

*RST reverts this setting to default value.

Syntax

:CONFigure[1..n]:ANAlysis:IORefraction<wsp>

<IOR>|MAXimum|MINimum|DEFault

Parameter(s)

IOR:
The program data syntax for <IOR> is defined as

a <numeric_value> element. The <IOR>

special forms MINimum, MAXimum and DEFault

are accepted on input.

MINimum allows to set the instrument to the

lowest supported value.

MAXimum allows to set the instrument to the

highest supported value.

DEFault allows the instrument to select a value

for the <IOR> parameter.

Sets the index of refraction.

Example(s)

CONF:ANA:IOR? Ex.: Returns 1.4677

CONF:ANA:IOR 1.5

CONF:ANA:IOR? Returns 1.5