Configure[1..n]:analysis: iorefraction – EXFO FTB-7000 OTDR Series for FTB-500 Benutzerhandbuch
Seite 411

SCPI-Befehlsreferenz
OTDR
397
Product-Specific Commands—Description
:CONFigure[1..n]:ANAlysis:
IORefraction?
Description
This query returns the index of refraction that will
be used for the next acquisition.
*RST reverts this setting to default value.
Syntax
:CONFigure[1..n]:ANAlysis:IORefraction?[<wsp
>MINimum|MAXimum|DEFault]
Parameter(s)
Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MINimum|MAXimum|DEFault.
MINimum is used to retrieve the instrument's
lowest supported value.
MAXimum is used to retrieve the instrument's
highest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax
<IOR>
Response(s)
IOR:
The response data syntax for <IOR> is defined
as a <NR3 NUMERIC RESPONSE DATA>
element.
Returns the index of refraction.
Example(s)
CONF:ANA:IOR 1.5
CONF:ANA:IOR? Returns 1.5