Ehe :fetch[1..n]:asetting:range, Fetch[1..n]:asetting:range – EXFO FTB-7000 OTDR Series for FTB-500 Benutzerhandbuch
Seite 424

SCPI-Befehlsreferenz
410
Serie FTB-7000
Product-Specific Commands—Description
:FETCh[1..n]:ASETting:RANGe?
Description
This query returns the range found after an
initiate (INIT) command. Note that acquisition
mode (CONF:ACQ:MODE) must be set to
ASETting.
Since *RST clears the range value, the returned
value will be 0.
Syntax
:FETCh[1..n]:ASETting:RANGe?
Parameter(s)
None
Response Syntax
<Range>
Response(s)
Range:
The response data syntax for <Range> is
defined as a <NR3 NUMERIC RESPONSE DATA>
element.
Returns the range, in meters.
Example(s)
CONF:ACQ:MODE ASET
INIT
INIT:STAT? Returns 0 when acquisition is
complete.
FETC:ASET:RANG? Ex.: Returns 1.25E+3
See Also
CONFigure[1..n]:ACQuisition:RANGe?
CONFigure[1..n]:ACQuisition:PULSe?
CONFigure[1..n]:ACQuisition:DURation?
CONFigure[1..n]:ACQuisition:WAVelength:LIST?
CONFigure[1..n]:ACQuisition:RANGe:LIST?
CONFigure[1..n]:ACQuisition:PULSe:LIST?
CONFigure[1..n]:ACQuisition