Configure[1..n]:analysis: rbscatter, Configure[1..n]:analysis:rbscatter – EXFO FTB-7000 OTDR Series for FTB-500 Benutzerhandbuch

Seite 412

Advertising
background image

SCPI-Befehlsreferenz

398

Serie FTB-7000

Product-Specific Commands—Description

:CONFigure[1..n]:ANAlysis:RBScatter

Description

This command sets the Rayleigh backscatter that

will be used for the next acquisition.

*RST reverts this setting to default value.

Syntax

:CONFigure[1..n]:ANAlysis:RBScatter<wsp><R

BS>|MAXimum|MINimum|DEFault

Parameter(s)

RBS:
The program data syntax for <RBS> is defined

as a <numeric_value> element. The <RBS>

special forms MINimum, MAXimum and DEFault

are accepted on input.

MINimum allows to set the instrument to the

lowest supported value.

MAXimum allows to set the instrument to the

highest supported value.

DEFault allows the instrument to select a value

for the <RBS> parameter.

Sets the Rayleigh backscatter.

Example(s)

CONF:ANA:RBS? Ex.: Returns –79.5

CONF:ANA:RBS –80

CONF:ANA:RBS? Returns –80