Product-Specific Commands—Description

:CONFigure[1..n]:ANAlysis:THReshold:

REFLectance

Description

This command sets the reflectance threshold

that will be used for the next acquisition.

*RST returns this setting to default value.

Syntax

:CONFigure[1..n]:ANAlysis:THReshold:REFLecta

nce<wsp><Reflectance>|MAXimum|MINimu

m|DEFault

Parameter(s)

Reflectance:
The program data syntax for <Reflectance> is

defined as a <numeric_value> element. The

<Reflectance> special forms MINimum,

MAXimum and DEFault are accepted on input.

MINimum allows to set the instrument to the

lowest supported value.

MAXimum allows to set the instrument to the

highest supported value.

DEFault allows the instrument to select a value

for the <Reflectance> parameter.

Sets the reflectance threshold.

Example(s)

CONF:ANA:THR:REFL? Ex.: Returns -72.0

CONF:ANA:THR:REFL -72.5

CONF:ANA:THR:REFL? Returns -72.5