Specifications, Specifications model t32.xs – WIKA T32.xS Benutzerhandbuch
Seite 14

GB
11258421.10 08/2014 GB/D
14
WIKA operating instructions model T32.xS
3. Specifications
Specifications
Model T32.xS
Permissible ambient temperature
range
{-50} -40 ... +85 °C
Climate class per IEC 654-1: 1993
Cx (-40 ... +85 °C, 5 ... 95 % r. h.)
Maximum permissible humidity
■
Model T32.1S
per IEC 60068-2-38: 1974
■
Model T32.3S
per IEC 60068-2-30: 2005
Test max. temperature variation 65 °C
and -10 °C, relative humidity 93 % ±3 %
Test max. temperature 55 °C, relative
humidity 95 %
Vibration per IEC 60068-2-6: 2007
Test Fc: 10 ... 2000 Hz; 10 g,
Amplitude 0.75 mm
Shock per IEC 68-2-27: 1987
Test Ea: Acceleration Type I 30 g and
Type II 100 g
Salt mist per IEC 60068-2-52
Severity level 1
Free fall drop per IEC 68-2-52: 1996 Drop height 1500 mm
Electromagnetic compatibility
(EMC)
1)
2004/108/EC, EN 61326 Emission
(Group 1, Class B) and immunity
(industrial application)
{ } Items in curved brackets are options for additional price, not for rail mounting version
T32.3S
1) During interference an increased measuring deviation of up to 1 % has to be considered.
For further specifications see WIKA data sheet TE 32.04 and the
order documentation.
For further important safety instructions for operation in
hazardous areas see chapter 10 “Notes for mounting and
operating in hazardous areas”.
3. Specifications