3. Specifications

Specifications

Model T32.xS

Permissible ambient temperature

range

{-50} -40 ... +85 °C

Climate class per IEC 654-1: 1993

Cx (-40 ... +85 °C, 5 ... 95 % r. h.)

Maximum permissible humidity

Model T32.1S

per IEC 60068-2-38: 1974

Model T32.3S

per IEC 60068-2-30: 2005

Test max. temperature variation 65 °C

and -10 °C, relative humidity 93 % ±3 %

Test max. temperature 55 °C, relative

humidity 95 %

Vibration per IEC 60068-2-6: 2007

Test Fc: 10 ... 2000 Hz; 10 g,

Amplitude 0.75 mm

Shock per IEC 68-2-27: 1987

Test Ea: Acceleration Type I 30 g and

Type II 100 g

Salt mist per IEC 60068-2-52

Severity level 1

Free fall drop per IEC 68-2-52: 1996 Drop height 1500 mm

Electromagnetic compatibility

(EMC)

1)

2004/108/EC, EN 61326 Emission

(Group 1, Class B) and immunity

(industrial application)

{ } Items in curved brackets are options for additional price, not for rail mounting version

T32.3S

1) During interference an increased measuring deviation of up to 1 % has to be considered.

For further specifications see WIKA data sheet TE 32.04 and the

order documentation.

For further important safety instructions for operation in

hazardous areas see chapter 10 “Notes for mounting and

operating in hazardous areas”.