Commissioning, operation – WIKA T32.xS Benutzerhandbuch
Seite 18

GB
11258421.10 08/2014 GB/D
18
WIKA operating instructions model T32.xS
6. Commissioning, operation
Mounting example:
3173801.A
6.2 Mounting
6.2.1 Transmitter in head mounting version (model T32.1S)
The transmitters for head mounting (model T32.1S) are designed to
be mounted on a measuring insert within a Form B, DIN connection
head, with extended mounting space. The connection wires of the
measuring insert must be approx. 50 mm long and insulated.
For applications with higher EMC requirements, it is recommended
using a shielded cable between the temperature transmitter and the
sensor, especially in connection with long leads to the sensor. For an
exemplary illustration, see drawing.
grounded
grounded shield
sensor wire
equipotential bonding
T32.xS
Field housing/
switch cabinet
For thermocouples, additionally ground here!
sensor
For grounding of the model T32.xS tempe-
rature transmitter see safety instructions
page 8f.