WIKA T32.xS Benutzerhandbuch

Seite 41

Advertising
background image

GB

11258421.10 08/2014 GB/D

WIKA operating instructions model T32.xS

41

10.3 Safety values

10.3.1 Sensor circuit (terminals 1 to 4)

Parameters

Model T32.xS.0IS,

T32.xS.0IS-x

Model T32.xS.0IC,

T32.xS.0IC-x

Level of protection

Ex ia IIC/IIB/IIA

Ex ia IIIC

Ex ic IIC/IIB/IIA

Terminals

1-4

Voltage U

o

DC 6.5 V

Strength of current l

o

9.3 mA

Power P

o

15.2 mW

Voltage U

o

n. a.

Strength of current l

o

n. a.

Power P

o

n. a.

Effective internal capacitance

C

i

208 nF

Effective internal inductance L

i

negligible

Max. external

capacitance C

o

IIC

24 µF

1)

325 µF

1)

IIB IIIC

570 µF

1)

570 µF

1)

IIA

1,000 µF

1)

1,000 µF

1)

Max. external

inductance L

o

IIC

365 mH

821 mH

IIB IIIC

1,644 mH

3,699 mH

IIA

3,288 mH

7,399 mH

Max. inductance/

resistance ratio

L

o

/R

o

IIC

1.44 mH/Ω

3.23 mH/Ω

IIB IIIC

5.75 µH/Ω

12.9 mH/Ω

IIA

11.5 µH/Ω

25.8 mH/Ω

Characteristics

linear

n. a. = not applicable

1) C

i

already incorporated

Notes:

U

o

: max. voltage of any conductor against the other three conductors

I

o

: max. current of three conductors parallel to the fourth conductor or any other combination

P

o

: max. power of three conductors parallel to the fourth conductor or any other combination

Due to separation requirements of the applied standards, IS supply- and

signal-circuit and the IS sensor circuit shall be considered as beeing

galvanically connected to each other.

10. Notes for mounting, operating in ...

Advertising