Commissioning, operation – WIKA T32.xS Benutzerhandbuch

Seite 19

Advertising
background image

GB

11258421.10 08/2014 GB/D

WIKA operating instructions model T32.xS

19

Mounting on the measuring insert

Mount the transmitter on the circular plate of the measuring insert

using two countersunk M3 screws per EN ISO 2009. Appropriate

threaded inserts have been press-fitted in the underside of the case.

Assuming the countersinking is carried out correctly, the permissible

screw length can be calculated as follows:

Check the screw length before fixing the transmitter to the measuring

insert: insert the screw into the circular plate and verify length of 4 mm!

CAUTION!

Do not exceed the maximum permissible screw length!

The transmitter will be damaged if the screws are

screwed further than 4 mm into the bottom of the

transmitter.

Mounting in connection head

Insert the measuring insert with the mounted transmitter into the

protective sheath and secure into the connecting head using screws

in pressure springs.

3173801.C

l

max.

= s + 4 mm

with

l

max.

Length of screw in mm

s Thickness of circular

plate in mm

3173801.B

max.

4

Circular plate

l

s

6. Commissioning, operation

Installation by means of DIN rail adapter

If the mechanical adapter, available as an accessory, is used the

T32.1S head transmitters can also be fixed on a DIN rail.

Advertising