Commissioning, operation – WIKA TR10-A Benutzerhandbuch
Seite 19

WIKA operating instructions RTD and TC, intrinsically safe designs (Ex i)
3345267.12 01/2014 GB/D
19
GB
6.3 Electrical connection of thermocouples
Cable colour coding of thermocouples
Sensor type
Standard
Positive
Negative
K
DIN EN 60584
Green
White
J
DIN EN 60584
Black
White
E
DIN EN 60584
Violet
White
T
DIN EN 60584
Brown
White
N
DIN EN 60584
Pink
White
6.3.1 Thermocouples with terminal block
Binder
Series 680
Binder
Series 680
Binder
Series 680
Binder
Series 680
Binder
Series 692
Binder screw/plug-in connector
3366142.02
Connector
(male)
View of the connector contacts
Socket (female)
View of the socket contacts
3166822.03
Single thermocouple
Dual thermocouple
The colour coding at the positive
connection to the devices always
decides the correlation of polarity
and connection terminal.
6. Commissioning, operation