IMPAC pyrometers ISR 12-LO/GS

21

9.6

Number of devices

With a click on “Number of devices” InfraWin changes to the display of 1 or 2 devices. If 2 devices
are selected, always 2 windows are displayed for settings or evaluation.



9.7

Basic settings

Under pyrometer parameters all preset values
can be displayed and modified if necessary. If the
window pyrometer parameters displayed, chang-
ing of settings on the pyrometer is blocked.

The window pyrometer parameter contains all parameter
settings described in section 8, Parameters.

Notes:

• “Basic range” displays the total range of the pyrometer

automatically and cannot be changed. If the sub range
is changed, confirm it with the small OK button which
appears.

• Under “Material “ you have the possibility to store the

names of different measuring objects with their emis-
sivity values and to recall them from the list.

• A click on the laser targeting light icon ( ) turns the

laser targeting light on or off at this point. After approx.
two minutes the laser targeting light is switched off
automatically.


The open / save button enable to store and recall own pyrometer configurations.

“1 meas.” shows the last pouring stream temperature in the pyrometer parameters window for
approx. 1 second.

If the pyrometer is used in another mode than pouring stream, „1 meas.“ shows the current
measuring temperature for approx. 1 second.

In this window the pouring stream mode can
be switched on and off (standard setting at the
ISR 12-LO/GS is “on”).
Descriptions to the parameters are to find un-
der 8.1 pouring stream mode).