1 dimensions, Dimensions – LumaSense Technologies ISR 12-LO/GS Benutzerhandbuch

Seite 6

Advertising
background image

IMPAC pyrometers ISR 12-LO/GS

6

Emissivity

ε

:

0.1 to 1 (adjustable in steps of 0.001)

Switch-off limit:

2% to 50% (in 1%-steps)

Exposure time t

90

:

2 ms (with dynamical adaptation at low signal levels),
adjustable to 0.01 s; 0.05 s; 0.25 s; 1 s; 3 s; 10 s

Maximum value storage:

Built-in single or double storage. Clearing with
adjusted time t

clear

(OFF; 0.01 s; 0.05 s; 0.25 s; 1 s; 5 s; 25 s)

, extern, via

interface or automatically with the next measuring object,

hold function to

freeze the actual displayed pouring stream temperature

Switch contact:

max. 0.15 A (only active with automatic clear mode or t

CL

≥ 0.5 s)

Measuring uncertainty:
(

ε

= 1, t

90

= 1 s, T

amb.

= 23°C)

0.4% of reading in °C + 1°C (< 1500 °C)
0.6% of reading in °C + 1°C (> 1500 °C)

Repeatability:

0.2% of reading in °C + 1°C

Control panel:

4 function keys, switch for serial interface, push button for test current

Protection class:

IP 65 (DIN 40 050)

Ambient temperature:

Converter housing: 0 to 60°C
Optical head: up to 250°C

Storage temperature:

-20 to 70°C

Weight converter:

2.2 kg

Sighting:

Laser targeting light (max. power level < 1 mW,

λ = 630-680 nm, CDRH class II). (measurement is
stopped while laser targeting light is switched on).

CE-label:

According to EU directives about electromagnetic immunity



1.1

Dimensions


Pyrometer housing:

Optical head:

cover


window

Lens
















all dimensions in mm