Programmable controller PI 6000
27
1.6
Disposal / decommissioning
Inoperable IMPAC controllers and pyrometers have to be disposed corresponding to the local regulations of
electro or electronic material.
2 Technical
data
Measured data display:
4-digit 7-segment LED display, 13 mm, red
Display of the unit of measure °C or °F, 7 mm, red
Program no. display:
Single digit 7-segment LED display, 13 mm, red
Controlled variable display:
10-section vertical LED bar display, approx. 4.5 mm wide, red
Controller cycle time:
250 µs
Proportional element Xp:
0.0...1000.0%
Integration time Ki:
0 to 10 minutes in 10 ms steps
Differential time Kd:
0 to 10 minutes in 10 ms steps
Controlled variable limiting:
0.0...100.0%
Duration of a program step
(segment):
Up to 27 minutes in
1
/
10
s steps, up to 4.5 hours in 1 s steps,
up to 45.5 hours in 10 s steps, "CONT" for continuous operation
Number of programs:
Maximum 9
Number of steps:
Maximum 20 per control program
AutoTune mode
For determining Xp and Ki (Kd = 0) for a control temperature
Power supply:
18... 30 V DC, ripple < 50 mV
pp
Power consumption:
Approx. 2.5 W (controller without external load)
Analogue input 1:
(measurement pyrometer)
0/4... 20 mA, input resistance approx. 120
Ω, for transferring the measured
values from the pyrometer to the PI 6000
Analogue input 2:
(optional alarm pyrometer)
0/4... 20 mA, input resistance approx. 120
Ω, for the connection of an addi-
tional pyrometer for shut-down in case of overtemperature (safety shut-
down). It is only possible to use alarm pyrometers with positive measuring
range limits.
Analogue output:
(electrically
isolated from the power supply)
0/4... 20 mA output, load max. 500
Ω, continuous controlled variable
0...100%
Digital interface to the py-
rometer:
Can be changed between RS232 or RS485 half duplex, for communication
with the measurement pyrometer. The transmission speed is automatically
adjusted to the same speed as is set on the PC interface.
Digital interface to the PC:
(electrically isolated from the power
supply)
RS232, transmission speed 9600... 38400 Bd
for connection to a PC or a higher level controller (PLC)
Control input:
Contact (normally open) for starting and stopping the selected control pro-
gram, power supply positive or negative can be used as the reference
Control outputs:
4 semiconductor relays, maximum load 50 VDC or 36 VAC 0.2 A
- Signal when a control program is running
- Signal when the generator is to provide power
- Signal when the alarm pyrometer (option) has triggered a safety shut-down
- Pulse of selectable length when the control program has been completed
Operating temperature:
0 ... 50°C at the housing
Storage temperature:
-20 ... 70°C
Weight:
Approx. 250 g
Ingress protection:
Front IP40, terminals IP20 (in accordance with DIN 40 050)
Protection class:
Protection class III (in accordance with EN61140)
Display resolution:
1°C or 1°F