Electrical connections on the rear of the unit

All electrical connections are on the rear of the unit. Con-
nectors with screw terminals are fitted to the male connec-
tor. The connectors can be removed and combined into
appropriate connection groups for easier installation.

The conductor cross-section that can be used in the con-
nectors is 0.14 ... 2.5 mm², with ferrules: 0.5... 1.5 mm².



5.3

Overview of connections

Connections

Function

1 / 2

Alarm pyrometer
4...20mA, 2-wire

2 / 3

Alarm pyrometer,
supply 4-wire *

)

4 / 5

Alarm pyrometer,
analogue input 4-wire

6 / 7 / 8

PC interface, RS232

9

Reference for 11 to 14

10 / 3

"start / stop program" trigger

11 / 9

"program start" switched output

12 / 9

"generator on" switched output

13 / 9

"ready" switched output
(pulse length can be selected)

14 / 9

"safety shut-down triggered"
switched output

15 / 16

Power supply PI 6000

17 / 18

Supply (measurement
pyrometer) *

)

19 / 20

Analogue input 0/4... 20 mA
(measurement pyrometer)

14

13

12

11

10

9

Tx

8

GND

7

Rx

6

-

5

+

4

-

3

+

2

-

1

+

27

-

26

25

24

GND

23

Tx

B

22

Rx

A

21

+

20

-

19

+

18

-

17

+

16

-

15

Input

Power supply

18...30 VDC

18...30VDC

output

0/4…20mA

RS232 / 485

Measurement pyrometer

RS232

controller output

0/4…20mA

Alarm pyrometer

(optional)

4…20mA

2-wire

18...30VDC

output

0/4…20mA

4-wire

to PC

RS232

safety shut-down (output)

ready (output)

generator on (output)

program start (output)

ext. trigger (input)

21 / 22 / 23 Interface (measurement pyrometer)
24 / 25

Jumper for setting RS232 to the measurement
pyrometer

26 / 27

Output, controlled variable 0/4... 20 mA

*

)

Looped through from 15 / 16



5.3.1 Power

supply

A voltage between 18 and 30 V DC is required to operate the PI 6000. The unit is
ready for use immediately if connected to the power supply. The unit does not have
an on/off switch; the power supply must be interrupted to switch off the unit.

Note: The power supply to the PI 6000 is also used as the supply for the
measurement pyrometer and the analogue alarm pyrometer, both for the
two-wire and the 4-wire version.