WIKA operating instructions model TR25 (Ex i)

14073045.01 10/2013 GB/D

19

GB

7.4 Mounting examples in hazardous areas
7.4.1 Possible installation methods with the marking II 1G Ex ia IIC T6 Ga or

II 1D Ex ia IIIC T65 °C Da

The sensor together with housing or connection head is located in Zone 0 (Zone 20). An Ex ia

type circuit must be used. Connection heads/cases made of aluminium are not permitted in

zone 0. At this position, WIKA uses connection heads/cases made of stainless steel.

Zones 0, 1, 2 or Zones 20, 21, 22

Hazardous area

Safe area

Thermowell

Thermowell

welded

TWxx

TWxx

Process

connection

Connection head/

Field housing

Tx10-B
Tx10-C

Tx10-C
Tx10-D

Tx10-H

Tx10-H

Option:

with built-in

transmitter:

e.g. T32

Tx10-A

Process

connection

Compression

fitting

Associated

electrical equipment

Intrinsically

safe supply or

suitable barrier

Intrinsically

safe supply or

suitable barrier

Intrinsically

safe supply or

suitable barrier

Intrinsically

safe supply or

suitable barrier

Tx10-A