10.3 Calibration, Recalibration

It is recommended that the measuring insert is recalibrated at regular intervals of

approx. 24 months. This period can reduce, depending on the particular application. The

calibration can be carried out by the manufacturer, as well as on site by qualified technical staff

with calibration instruments.

11. Faults

Faults

Causes

Measures

No signal/

line break

Mechanical load too high or

overtemperature

Replacement of the resistance thermometer

Erroneous measured

values

Sensor drift caused by

overtemperature

Replacement of the resistance thermometer

Erroneous measured

values (too low)

Entry of moisture into cable

Replacement of the resistance thermometer

Erroneous measured

values and response

times too long

Wrong mounting geometry or heat

dissipation too high

The temperature-sensitive area of the

sensor must be inside the medium. Beware

of the mounting position of the sensor!

Deposits within the thermowell

Remove deposits

Indication of the

measured value jumps

Cable break in connecting cable

or loose contact caused by

mechanical overload

Check the supply cable

Corrosion

Composition of the medium not

as expected or modified or wrong

thermowell material selected

Analyse medium and then select a more-

suitable material or replace thermowell

regularly

Signal interference

Stray currents caused by electric

fields or earth loops

Use of screened connecting cables,

increase in the distance to motors and

power lines

Earth circuits

Elimination of potentials, use of galvanically

isolated transmitter supply isolators or

transmitters

CAUTION!

If faults cannot be eliminated by means of the measures listed above, shut down the

instrument immediately, and ensure that pressure and/or signal are no longer present,

and secure the instrument from being put back into operation inadvertently.

In this case, contact the manufacturer.

If a return is needed, follow the instructions given in chapter 12.2 “Return”.