6. Object manager

7.2 Setup of the model CPU6000-S

1. Connect the CPU6000-S to the PC via the supplied USB cable.

2. Directly download the driver from the local website and install.

http://en-co.wika.de/cpu6000_w_cpu6000_s_cpu6000_m_en_co.WIKA

→ Software → Interface driver

The driver has been correctly installed when, in Windows Device

Manager, by switching the CPU on or off, a COM port appears under

“Connections”.

3. In WIKA-CAL, within “Object manager” under “Workplace”, select the new COM

port under “COM-Port” of the definded pressure balance and save the data record.

The pressure balance must first be generated and assigned to the work station (see

chapter 6.3.18 “References (Pressure)”).

4. Set baud rate 9600 and, via the [Test] button, check the communication. The

communication is working properly if a realistic value is displayed.

5. In “Object manager”, under “Piston”, the “Floating position” for the respective

piston of the pressure balance must be specified. This can be entered manually in

[mm] or read via the [Read position] button. The data record must then be saved

with [Save] in order to accept the settings.