6. Object manager

7.3 Setup of the model CPU6000-M

1. Connect the CPU6000-M to the PC via the supplied USB cable.

2. Directly download the driver from the local website and install.

http://en-co.wika.de/cpu6000_w_cpu6000_s_cpu6000_m_en_co.WIKA

→ Software → Interface driver

The driver has been correctly installed when, in Windows Device

Manager, by switching the CPU on or off, a COM port appears under

“Connections”.

3. In WIKA-CAL, within “Object manager” under “Workplace”, select the new COM

port under the “COM-Port” of the defined CPU6000-M (multimeter) and save the

data record. The multimeter must first be generated and the work station assigned

(see chapter 6.3.17 “Multimeter”).

4. Set baud rate 9600 and, via the [Test] button, check the communication. The

communication is working properly if a realistic value is displayed.