IMPAC pyrometers ISQ 5 · ISQ 5-LO
10
Pyrometer
230V ~
24 V DC
°C
yell ow
LED digital
displa y
controller
po wer sup ply
green
bro wn
white
printer
3.5
Connection schematic for analyzing devices
Additional analyzing instruments, for
example a LED digital display instru-
ment only needs to be connected to a
power supply and the analog outputs
from the pyrometer (exception: the
digital display DA 6000 also can be
connected with its serial interface, the
digital display DA 6000-N has to be connected with its serial interface).
Another Instruments like a controller or printer can be connected to the
display in series as shown above (total load of resistance max. 500
Ohm).
4
Mechanical installation
4.1
Accessories (option)
Numerous accessories guarantee easy installation of the pyrometers. The following overview shows a selec-
tion of suitable accessories. You can find the entire accessory program with all reference numbers on section
14.2, Reference numbers.
Mounting (for ISQ 5):
For easy mounting and aligning the pyrometer to the
measured object an adjustable mounting angle is avail-
able.
Mounting angle
Cooling (for ISQ 5):
The completely covered water cooling jacket made from
stainless steel protects the pyrometer if exposed to a hot
environment. It is designed for ambient temperatures up
to 180°C.
Water cooling jacket with
integrated air purge
Miscellaneous (for ISQ 5):
The air purge protects the lens from contamination with
dust and moisture. It has to be supplied with dry and oil-
free pressurized air and generates an air stream shaped
like a cone.
The pyrometer can be easily fixed on a vacuum chamber
with the KF 16 vacuum support with sighting window.
Air purge
Vacuum support
The scanning mirror unit SCA 5 (for ISQ 5): enables the meas-
ured object to be scanned over a certain range. The measuring
beam of the pyrometer moves straight in one line across the object
and collect temperature data of this line. This is useful in combina-
tion with the maximum value storage (peak picker) to measure ob-
jects which move out of the target area. The scanning angle of the
mirror is 0 ... 12°, the scanning frequency 0 ... 5 Hz. Both values are
easily adjustable at the instrument.
The mounting tube (for ISQ 5):is used for fixing of the pyrometer at
machines and protects the instrument. It is equipped with an air
connector to cool the pyrometer with compressed air and to purge
the optics to keep it clean.
The flange tube is similar to the mounting tube, it will be delivered
with an additional flange for fixing. This flange has to be welded in
the required position.