LumaSense Technologies ISQ 5-LO Benutzerhandbuch
Seite 27

IMPAC pyrometers ISQ 5 · ISQ 5-LO
27
This is followed, if necessary for that command, by the ASCII parameter “X”. If this parameter “X” is omitted,
then the device resets with the current parameter.
A “?” after the small command letters answers with the respective settings (only at setting commands, not at
enquiry commands).
stand for product of emissivity, surface coverage and transmission of the measuring distance
Example: Entry: “00em“ +
CR
The emissivity setting (
) of the device with the address 00 is returned
Answer: “0970“ +
CR means Emissivity = 0.97 or 97.0%
Description
Command
Parameters
1
/
2
write
read
AAevXXXX
AAvr
XXXX=0800..1250
1
/
2
=0.800..1.250
Answer: DDDD 4 decimal digits 0800..1250
for one-channel
temperature
AAemXXXX
AAem
XXXX=0050..1000
=0.050...1.000
Answer: DDDD 4 decimal digits 0050...1000
Response time
AAezX
X=0..6 0=0.00s
1=0.01s 2=0.05s 3=0.25s
4=1.00s
5=3.00s
6=9.99s
Clear peak memory
AAlzX
X=0..8 0=
OFF
1=0.01s 2=0.05s 3=0.25s
4=1.0s
5=5.0s
6=25.0s
7=EXTERN
8=AUTO
External clearance
AAlx
Analog output
AAasX
X=0 (0…20mA) X=1 (4…20mA)
Laser / One-channel
mode
AAlaX
AAla
X=0 off
X=1 on
Answer:
1 digit 0 / 1
Measuring value
(ratio temperature)
AAms
Answer: QQQQQ (88880=Overflow)
5 decimal digit (in °C, last digit is 1/10 °C)
Measuring value
(one-channel and
ratio temperature)
AAek
Answer: SSSSSQQQQQ
2x5 decimal digits (in °C, last digit is 1/10°C), one-channel temperature with
emissivity
(read)
AAtr
Answer: DDDD
4 decimal digit 0000..1500
Minimum intensity
(minimal allowed
)
AAawXX
AAar
XX=02…50
(0.020..0.500)
Answer: DD
2 decimal digit 02..05
Basic range (read)
AAmb
Answer: XXXXYYYY 2x4 hex-digit for lower and upper range limit (°C)
Sub range (read)
AAme
Answer: XXXXYYYY 2x4 hex-digit for lower and upper range limit (°C)
Sub range (set) 1.
2.
AAm1XXXXYYYY
AAm2
XXXXYYYY=2x4 hex-digit for lower and
upper range limit (°C)
AAm2 confirms the change
(autoreset)
internal temperature
(read)
AAgt
AAtm
Answer: DD
2 decimal digits (00..98 °C)
gt=current temp. tm=maximum temp. (memory)
Baud rate (set)
AAbrX
X=0..5 0=1200 Bd ... 5=38.4 kBd (autoreset)
Device address (set) AAgaXX
XX=00...97 (autoreset)
Read parameters
AApa
Answer: 15 decimal digits
DD............. : Emissivity
(see em)
..D............ : Response time
(see ez)
...D........... : Clear peak memory (see lz)
....D.......... : Analog output
(see as)
.....DD........ : Internal temperature (see gt)
.......DD...... : Device address
(see ga)
.........4..... : Baud rate
(see br )
..........0.... : always 0
...........DDDD : Ratio correction
(see vr)
Device type /
software version
AAve
Answer: VVMMJJ VV=54
MM=Month JJ=Year of software version
Extended data protocol for video module
Reading video-status:
AAos
Output: XX hexadecimal-Byte
(bits 6, 5 and 3 unassigned)
Bit 7 = 1
no watch/date implemented
Bit 4 = 1
date/watch had a low voltage error
Bit 2 = 1
date is indicated
Bit 1 = 1
watch is indicated
Bit 0 = 1
user text
= 0 device number is indicated
User-text:
AAox
AAox_
AAoxTT…TT
Output: "XXXXXXXXXXXX"
ASCII-symbol with "" limited
_ = space character deletes bit 0 in video-status (auto reset!)
TT...TT user text with max. 12 ASCII, sets bit 0 in video-status (auto
reset!)