1 dimensions, Dimensions – LumaSense Technologies ISQ 5-LO Benutzerhandbuch
Seite 6

IMPAC pyrometers ISQ 5 · ISQ 5-LO
6
0.1°C (at 900°C measuring temperature)
MB 30: 0.2°C (at 1010°C measuring temperature)
0.1°C (at 1100°C measuring temperature)
Temp. dependence:
0.25°C per °C deviation of ambient temperature from 25°C
Sighting system:
ISQ 5-LO: Laser targeting light (max. power level < 1 mW,
= 630-680 nm, CDRH class II)
ISQ 5:
Laser targeting light (max. power level < 1 mW,
= 630-680 nm, CDRH class II)
or thru-lens view finder or video module
Protection class:
IP65 (DIN 40050)
Ambient temperature:
0 to 70°C
Storage temperature:
-20 to 70°C
Weight:
550 g
Housing:
stainless steel
CE-label:
according to EU directives about electromagnetic immunity
Additional technical data for pyrometers with built-in video module (ISQ 5-TV):
Video signal:
BAS signal approx. 1 V
pp
at 75
,
CCIR, 50 Hz
NTSC-Norm, EIA, 60 Hz
(optional to order)
Array size:
628 x 583 pixels, black & white
510 x 492 pixels, black & white
Exposure:
automatic, additionally 3-levels controlled by the measuring temperature
Field of view:
approx. 10% x 14% of focused distance
Date/time:
real-time clock with at least 3 days spare run
Video output plug:
separate round plug at the pyrometer, not galvanically separated to the
pyrometer’s power supply
Picture insertions:
target marking; unit number or user text (max. 12 characters), time and/or date;
(individually switchable), measuring temperature, emissivity slope or emissivity
1.1
Dimensions
Optical head I:
Optical head II (fixed adjusted):
Optical head II (focusable):
All dimensions in mm