Commissioning, operation – WIKA T12 Benutzerhandbuch

Seite 16

Advertising
background image

GB

3211517.07

07/2010

GB/D

16

WIKA operating instructions temperature transmitter T12

Mounting example:

3173801.A

Mounting on the measuring insert

Mount the transmitter on the circular plate of the measuring insert

using two countersunk M3 screws per DIN EN ISO 2009. Appropriate

threaded inserts have been press-fitted in the underside of the case.

Assuming the countersinking is carried out correctly, the permissible

screw length can be calculated as follows:

l

max.

= s + 4 mm

with

l

max.

Length of screw in mm

s Thickness of circular

plate in mm

3173801.B

max.

4

Circular plate

l

s

6. Commissioning, operation

6. Commissioning, operation

In hazardous areas, only use temperature transmitters

that are approved for those hazardous areas. The

approval is marked on the product label.

6.1 Mounting

6.1.1 Transmitter in head mounting version (model T12.10)

The transmitters for head mounting are designed to be mounted

on a measuring insert within a Form B, DIN connection head, with

extended mounting space. The connection wires of the measuring

insert must be approx. 50 mm long and insulated.