Commissioning, operation, Model screwdriver tightening torque – WIKA T12 Benutzerhandbuch

Seite 19

Advertising
background image

GB

3211517.07

07/2010

GB/D

WIKA operating instructions temperature transmitter T12

19

Recommended tools for terminal screws:

Model

Screwdriver

Tightening torque

T12.10

Cross head ('Pozidriv' tip)

size 2 (ISO 8764)

0.4 Nm

T12.30

Slotted, 3 mm x 0.5 mm

(ISO 2380)

0.4 Nm

6.2.1 Power supply / 4 ... 20 mA loop

The T32 is a 2-wire temperature transmitter. Depending on the

version, it can be supplied with various types of power supply.

Connect the positive line of the power supply to the terminal marked

with ⊕, the negative line of the power supply to the terminal marked

with ⊖.

With flexible leads we recommend the use of crimped connector

sleeves.

The model T12 temperature transmitter requires a minimum terminal

voltage of DC 9 V. The load must not be too high, as otherwise, in the

case of relatively high currents, the terminal voltage at the trans-mitter

will be too low.

Maximum permissible load depending on the supply voltage:
Load diagram

6. Commissioning, operation

11289130.01

Voltage U

B

in V

Ex ia Ex nA/nL/ic

1174

913
652

0

0

9

24

30

36

Load

RA

in

Advertising