6 disposal / decommissioning, 2 technical data, Disposal / decommissioning – LumaSense Technologies ISR 50-LO Benutzerhandbuch

Seite 33: Technical data, 2technical data

Advertising
background image

INFRATHERM pyrometer ISR 50-LO

33

1.6

Disposal / decommissioning

Inoperable IMPAC pyrometers have to be disposed corresponding to the local regulations of electro or elec-
tronic material.

2

Technical data

Temperature ranges:

700 ... 1800°C

(MB 18)

800 ... 2500°C

(MB 25)

1000 ... 3000°C

(MB 30)

Sub range:

any range adjustable within the temperature range, minimum span 51°C

Spectral ranges:

channel 1: 0.9 µm
channel 2: 1.05 µm

IR detector:

Silicon foto diode (Si/Si)

Fibre:

MB 18: HD multi fibre 0.6 mm (green fibre mark)
MB 25: HD mono fibre 0.2 mm (red fibre mark)
MB 30: HD mono fibre 0.1 mm (yellow fibre mark)

Power supply:

24 V DC (18 ... 36 V DC), ripple < 500 mV

Power consumption:

Max. 1 W

Analog output:

0 ... 20 mA or 4 ... 20 mA (linear), switchable;
Test current 10 mA or 12 mA by pressing test key

Load:

0 ... 500

Ω

Digital Interface:

RS232 or RS485 addressable (half duplex), switchable;
baud rate 1200 up to 115200 Bd

Resolution:

0.1°C on interface and display;
< 0.1% of temperature range at the analog output

Isolation:

power supply, analog output and digital interface are
galvanically isolated from each other

Internal LC display:

LC display for temperature indication or parameter settings

Parameters:

Adjustable or readable at the instrument or via interface:

Measuring temperature, operation mode (ratio / mono), emissivity slope K
or emissivity

ε, exposure time t

90

, clear times t

clear

for maximum value

storage incl. automatically or external deletion of maximum value storage,
or hold function, analog output 0 ... 20 or 4 ... 20 mA, sub range, switch-
off level, contamination limit, RS485 address, baud rate, RS485-wait time
t

W

, temperature display in °C or °F, error status, maximum

internal temperature

Emissivity slope K:

0.8 ... 1.2 adjustable in steps of 0.001

Emissivity

ε

:

5 ... 100% adjustable in steps of 0.1%

Switch-off level:

2% ... 50%, adjustable

Exposure time t

90

:

10 ms; adjustable to 0.01 s; 0.05 s; 0.25 s; 1 s; 3 s; 10 s

Maximum value storage:

Built-in single or double storage. Clearing with adjusted time t

clear

(off; 0.01 s;

0.05 s; 0.25 s; 1 s; 5 s; 25 s), extern, via interface or automatically with the
next measuring object

Switch contact
Opto relay (AC/DC ):

Switch contact for dirty window alarm
max. switch current 0.5 A
max. switch supply 60 V AC/DC

Uncertainty:

(

ε = 1, t

90

= 1 s, T

amb.

= 23°C)

up to 1500°C:

0.5% of measured value in °C + 2°C

above 1500°C: 1.0% of measured value in °C

Repeatability:

(

ε = 1, t

90

= 1 s, T

amb.

= 23°C)

0.2% of measured value in °C + 2°C

Ambient temperature:

0 ... 50°C at the converter
0 ... 250°C at optical head

Storage temperature:

-20 ... 60°C