10 wait time (tw rs485), 11 temperature display (c / f), 12 error status (status) – LumaSense Technologies ISR 50-LO Benutzerhandbuch

Seite 45: 13 maximum internal temperature (maxinttemp), 10 settings via interface and software, 1 installation, 2 program start, Wait time (tw rs485), Temperature display (c / f), Error status (status)

Advertising
background image

INFRATHERM pyrometer ISR 50-LO

45

9.10

Wait time (tw RS485)

Using a pyrometer with RS485 it is possible that the connection is not fast enough to
receive the pyrometer’s answer to a command of the master. In this case a minimum
delay time (tw) can be set. The pyrometer waits this time until it answers a master in-
quiry at the RS485 interface (e.g.: tw = 02 at a baud rate 9600 means a wait time of

2

/

9600

sec).

Note: the setting of a delay time (tw) does not guarantee an answer to some commands directly after this
time. Certain commands require an internal operation time of max. 3 ms.


9.11

Temperature display (C / F)

The temperature can be displayed in °C or °F.


9.12

Error status (Status)

In case of a device error the pyrometer displays „Error“ including a hex code which identifies this error to
IMPAC service. The standard display at this point is “ok”.


9.13

Maximum internal temperature (MaxIntTemp)

Shows the maximum internal temperature the device ever reached.

10

Settings via interface and software


The operating and analyzing software InfraWin is included in delivery of the pyrometer. With this software all
pyrometer functions also can be used on the PC (except changing the interface or using the test current
function). Additionally the measuring values can be displayed and analyzed graphically and numerically.
This section gives an overview about the functions of the software. Additionally there is a description of the
individual icons in the program’s help menu. Click on the F1 button after InfraWin has been loaded or click
on the ? in the menu bar.

The following descriptions refer to program version 4.0. The latest version is available for free as download
from the homepage www.impacinfrared.com.


10.1

Installation

For installation select the setup program „setup.exe“ from the InfraWin-CD or from the downloaded and
unpacked zip file from the internet and follow the installation instructions.


10.2

Program start

After installation and the first program start a language must be chosen (German, English, French, Italian,
Spanish. The language also can be changed in the program at any time). On the start page the screen
shows the following icons:













Settings:

00 Bit

.

.

.

99 Bit

Settings:

°C

°F