2 thermal alignment, 8 instrument settings, 1 settings at the instrument – LumaSense Technologies ISR 50-LO Benutzerhandbuch

Seite 40: Thermal alignment, Instrument settings, Settings at the instrument, 8instrument settings

Advertising
background image

INFRATHERM pyrometer ISR 50-LO

40

7.2

Thermal alignment

When measuring a hot object in font of a colder background, it is usually sufficient to align the optical head
“thermally” to achieve the highest temperature (output) signal. For this kind of alignment the pyrometer must
be switched into 1-color mode, because in ratio mode the maximum signal can already be achieved even if
the object fills the spot only partially.


7.3

Avoiding measuring errors caused by incorrect installation

To avoid measuring errors, please note the following points when mounting the pyrometer:
• Infrared radiation of hot machine parts can be reflected from the measuring object and will be detected

additionally by the pyrometer. For example if the measured object has a high reflectivity, the measured
temperature will be mainly from the reflexions of the hot machine parts, not only from the measuring ob-
ject itself. In such a case the reflected radiation has to be blocked by suitable equipment.

• The spectral range of the ISR 50-LO is sensitive to bulb light or daylight (not for neon lamp light or dif-

fuse daylight). The influence of such external light must be avoided for an accurate measurement.


8

Instrument settings


The pyrometers are equipped with a wide range of settings for optimal adaptation to the required measuring
condition and for getting the correct measuring temperature.
All instrument settings can be done directly at the instrument or via serial interface and software InfraWin
(see section 10, Settings via interface and software, user of an own communication software find all inter-
face commands on section 14, Data format UPP

®

).



8.1

Settings at the instrument

The LC-display as well as the push buttons for
displaying and setting of the parameters are
found inside the converter. The pyrometer is
opened by 4 allen screws.

Note:

Only operate the device for a prolonged period of time when it is closed! After setting
the parameters, replace the housing cover immediately to prevent dirt from entering
the device and to meet the electromagnetic requirements (EMV).

1 PAR: With

the

PAR button all available parame-

ters are displayed in the following descrip-
tion (section 9). Pushing the button again
changes the display to the next parameter
and on the display a corresponding short
form is displayed.

2

:

With the arrow keys all parameter settings
can be displayed. They are active after
pushing the PAR button. Pushing the button
longer changes the settings in fast mode.

Push button
short description

Temperature
display or para-
meter short form

1

3

2

Push buttons

3 ESC / ENT: Pushing the ESC button changes the pyrometer to measuring mode. If a parameter is

changed with the arrow keys the indication of the ESC button changes to ENT. Pressing
the button again confirms the value into the pyrometer. Changing the parameters again by
pushing the PAR button doesn’t confirm this value in the pyrometer. If no button is pressed
for 30 s the pyrometer changes to the temperature indication without accepting the
changed value.

Note:

If the pyrometer is connected to a PC via interface and the software InfraWin is
opened with the setting window „Pyrometer Parameters“, changing of settings on the
pyrometer is blocked.


Interface switch

Push buttons

Test current key